Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-01595-1 |
Product Name/Description |
Philips Ingenia 1.5T and Ingenia 3.0T MR systems with Ingenia MR-OR solution
ARTG Number: 230064 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
12/12/2016 |
Responsible Entity |
|
Reason/Issue |
There is an electrical component called MR-OR Laser control board in the FlexTrak OR trolley, under the Interface Adapter. In production, two MR-OR Laser control boards have been found where the Laser control board has been assembled incorrectly, leading to insufficient isolation distance. Hence, there is a possibility that a patient or user may be exposed to a very small, level of touch current. In practice, the current is so low that it may not even be noticed. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Philips has revised the design of the control board and is sending out service engineers to install the revised boards. This action has been closed-out on 23/03/2018. |
Contact Information |
1800 251 400 - Philips Customer Care Centre |