Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-01427-1
Product Name/Description VITROS Chemistry Products C-reactive protein (CRP) Slides

Product Codes: 1926740, 8097990

Unique Identification Numbers: 10758750000296, 10758750004355

Affected Generations (GENs): 3, 11, 12, 30 through 36

This issue may affect expired, current (in date) and future product GENs

ARTG Number: 232251
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 4/11/2016
Responsible Entity Ortho-Clinical Diagnostics Australia Pty Ltd
Reason/Issue Ortho Clinical Diagnostics (Ortho) received reports of positively biased CRP results using VITROS CRP Slides when diluting samples. Investigation undertaken by Ortho identified that samples diluted (3X recommended dilution or higher) using VITROS Chemistry Products Specialty Diluent, VITROS Chemistry Products FS Diluent Pack 3 or a low CRP concentration sample had positively biased results if the endogenous level of CRP in the diluent is not subtracted from the final result. The magnitude of bias increases as the dilution factor increases.
Recall Action Recall for Product Correction
Recall Action Instructions Users are advised to follow the instructions in the customer letter to perform dilutions when processing samples above the measuring range for CRP. VITROS CRP Slides Instructions for Use (IFUs) will be revised to include the additional instructions to follow when diluting samples with results above the measuring range. Discuss any concerns regarding previously reported CRP results with the Laboratory Medical Director to determine the appropriate course of action. The results from this or any other diagnostic test should be used and interpreted only within the context of the overall clinical picture. This action has been closed-out on 26/05/2017.
Contact Information 1800 032 359 - OrthoCare Technical Solutions Centre