Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-01425-1
Product Name/Description Freelite Human Lambda Free kit for use on SPAPLUS

Item Numbers: TBSLK018S, TBSLK01810S

Lot Numbers: 400012, 400257

Expiry: Nov 2017
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 9/11/2016
Responsible Entity In Vitro Technologies Pty Ltd
Reason/Issue The manufacturer has advised that investigations of the products have identified a change in the calibration curve shape that has occurred since product release. Increased activity in the calibration curve between calibrators 5 and 6 has been observed and they have received reports of increased false prozone flags which is consistent with this issue.
Recall Action Recall
Recall Action Instructions Customers are advised to immediately quarantine and dispose of any affected lots held by the customers and also to remove any affected lots from the SPAPLUS instrument. It is also advised to discuss the issues with the Medical Director to identify any further actions are required including reviewing the previously reported results.

This action has been closed out on 30/10/2018
Contact Information 1300 552 003 - In Vitro Technologies