Type of Product Medical Device
TGA Recall Reference RC-2016-RN-01413-1
Product Name/Description Spot-light Tissue Pre-treatment Kit (Antigen retrieval reagent used in cellular imaging). An in vitro diagnostic medical device (IVD).

Catalogue Number: 008401

Lot Numbers: 1667961A (exp. 31 Jul 2016), 1713166A (exp. 30 Nov 2016), 1714439A (exp. 30 Apr 2017), 1763308A (exp. 28 Feb 2018)

ARTG Number: 212745
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 1/11/2016
Responsible Entity Life Technologies Australia Pty Ltd
Reason/Issue It has been identified that the enzyme digestion reagent (Reagent 2) contains 1.3X the normal amount of enzyme (pepsin) leading to excessive pepsinisation of cells. This may cause loss of tumour cell morphology and inadequate hybridization of FISH/CISH probes. The over pepsinisation will be obvious to the user when examining slides under a microscope.
Recall Action Recall
Recall Action Instructions Life Technologies is advising users to dispose of affected stock. Credit will be issued for the affected stock. Unaffected stock is expected to be available from 31 November 2016. This action has been closed-out on 18/05/2017.
Contact Information 1800 636 327 - Life Technologies Australia