Type of Product Medical Device
TGA Recall Reference RC-2016-RN-01335-1
Product Name/Description Blood Pressure Monitoring Kits with Transpac IV Transducers

Kit - Royal Women Hospital QLD, Special Care Nursery (Item: 011-0J069-01, Lot: 3253136)
Kit - Neonatal/Childrens Australia (Item: 011-46103-16, Lot: 3263780)
Safeset Transpac IV (Item: 011-46103-25, Lot: 3207778)
Transpac IV (Item: 011-46104-49, Lot: 3242280)
Single Line Monitoring Kit w/Transpac IV (Item: 011-46104-52, Lot: 3253165)
Single Line Transpac IV Monitoring Kit (Item: 011-46107-16, Lot: 3235355)

ARTG Number: 195624
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 19/10/2016
Responsible Entity ICU Medical Australia Pty Ltd
Reason/Issue It has been identified that there is a potential for leaks with certain monitoring kits that contain the Transpac IV transducers. A small number of these devices may have the potential to leak or introduce small amounts of air into the line. In a worst case, there is potential patient harm due to air embolism or contamination. ICU Medical, Inc. has not received any reports of adverse events associated with this issue.
Recall Action Recall
Recall Action Instructions ICU Medical is advising customers to arrange for their stock to be recovered and replacement stock, or a credit note to be issued. This action has been closed-out on 24/05/2017.
Contact Information 02 9629 6977 - ICU Medical