Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-01323-1
Product Name/Description ALBAclone Advanced Partial RhD Typing Kit. An in vitro diagnostic medical device (IVD).

Product Code: Z293

Affected Lot Numbers:
V163020, expiry 11 Aug 2017
V157640, expiry 9 Mar 2017

ARTG Number: 225063
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 13/10/2016
Responsible Entity Haemokinesis Pty Ltd
Reason/Issue The shelf life of two lots of the ALBAclone Advanced Partial RhD Typing Kit has been reduced from 24 months to 12 months based on stability testing results for component L (Anti-D Cell Line LHM57/17). The revised shelf life for each of these lots of kit has now expired and the kits are being recalled.
Recall Action Recall
Recall Action Instructions Haemokinesis is advising users to quarantine and dispose of the affected lots. Haemokinesis will provide replacement stock or will issue a credit note. For previously generated results, the laboratory Medical Director should take into consideration the possible reduced reactivity of component L.

This action has been closed out on 31/10/2018