Type of Product Medical Device
TGA Recall Reference RC-2016-RN-01264-1
Product Name/Description Radiometer TCM5 Flex Monitor (non-invasive blood gas monitor)

Serial Numbers: below 393-500R0030N009

ARTG Number: 140333
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 23/09/2016
Responsible Entity Radiometer Pacific Pty Ltd
Reason/Issue Radiometer has recently become aware of a potential risk that the TCM5 monitor may not be compliant with the waterproof specification, IPX2, which is required for a homecare environment. The problem is caused by incorrect assembly and may allow water to enter into the monitor, which may pose a risk of electrical shock for the user. There have been no reports of injuries as a result of this issue.
Recall Action Recall for Product Correction
Recall Action Instructions Radiometer is advising users to stop use of the affected devices until they have been inspected and repaired where required by a Radiometer representative. A loan unit will be provided while the device is being repaired. This action has been closed-out on 17/05/2017.
Contact Information 1800 247 254 - Radiometer