Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-01254-1
Product Name/Description VITROS Automation Solutions configured with a Thermo Scientific Centrifuge Module. An in vitro diagnostic medical device (IVD)

VITROS Automation Solutions with TCAutomation System (TAC)
Product Code: 6844300
Software version: 3.6.2 and below

Thermo Scientific Centrifuge Module
Product Code: 952040

ARTG Number: 229946
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 23/09/2016
Responsible Entity Ortho-Clinical Diagnostics
Reason/Issue Ortho has become aware that there is the potential for test results to be mis-associated with the wrong Sample ID (SID) and to be reported prior to the TCAutomation System detecting a SID mismatch. This mis-match may occur when the Centrifuge Module fails to grip the sample tube when attempting to unload the centrifuge.

The system will generate an error message for the unload failure, however the centrifuge will continue to operate as normal. The sample rack containing the sample that has not been unloaded will be transferred back to the load side of the centrifuge, and the system will not recognise that the sample is still in the rack. This may result in the sample being re-analysed and the SID being mis-associated.

Assay results are potentially associated with the wrong patient and may be reported. This could lead to misdiagnosis or inappropriate treatment of a patient. Ortho received no reports of patient injury due to this issue.
Recall Action Recall for Product Correction
Recall Action Instructions Ortho is advising users to check samples associated with Cross Check errors when they appear on the TAC error messages screen, and to ensure preventative maintenance and cleaning procedures for the centrifuge are carried out as required.

Past occurrences of this issue are not easily identifiable, and a review of previous results may be impractical. Ortho is recommending users discuss any concerns regarding previously reported results with the Laboratory Medical Director to determine the appropriate course of action. This action has been closed-out on 18/05/2017.
Contact Information 1800 032 359 - Ortho Care Technical Solutions Centre