Type of Product Medical Device
TGA Recall Reference RC-2016-RN-01216-1
Product Name/Description Getinge ED-FLOW AER, ED-FLOW SD AER (Automated washer-disinfector for endoscopes)

Model Number: 01280010

Serial Numbers: 4H120383, 4H120384

ARTG Number: 228574
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 15/09/2016
Responsible Entity Getinge Australia Pty Ltd
Reason/Issue The manufacturer has become aware that due to a software issue, if the compressed air supply of the machine is lost right before or during product supply (DLC detergent, Aperlan Poka-Yoke Agent A/B disinfectant), and after cycle start, and not re-established before cycle end, the machine does not launch or complete the product supply when reprocessing a scope. As a result, no alarm is triggered.
Recall Action Recall for Product Correction
Recall Action Instructions Getinge is advising users of interim instructions to follow. Users are advised to carefully examine the printer ticket at the end of each reprocessing cycle to ensure there are 3 lines printed for DLC weight, Disinfectant A and B weights are OK. A software upgrade will be implemented as a permanent correction. This action has been closed-out on 20/02/2017.
Contact Information 1300 155 500 - Getinge Australia