Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-01209-1 |
Product Name/Description |
AQUIOS CL Flow Cytometer System
Part Number: B30166
ARTG Number: 208574 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
14/09/2016 |
Responsible Entity |
|
Reason/Issue |
As a result of a software error, the use of Positions 9 and 10 of the reagent carousel can lead to misalignment of the carousel and damage to the prep probe. Damage to the prep probe will render the instrument inoperative with the potential to delay reporting of results. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Beckman Coulter recommends that positions 9 and 10 of the reagent carousel not be utilised until further notice.
This action has been closed out on the 25/09/2018. |
Contact Information |
1800 060 881 - Beckman Coulter Customer Support Centre |