Type of Product Medical Device
TGA Recall Reference RC-2016-RN-01209-1
Product Name/Description AQUIOS CL Flow Cytometer System

Part Number: B30166

ARTG Number: 208574
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 14/09/2016
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue As a result of a software error, the use of Positions 9 and 10 of the reagent carousel can lead to misalignment of the carousel and damage to the prep probe.
Damage to the prep probe will render the instrument inoperative with the potential to delay reporting of results.
Recall Action Recall for Product Correction
Recall Action Instructions Beckman Coulter recommends that positions 9 and 10 of the reagent carousel not be utilised until further notice.

This action has been closed out on the 25/09/2018.
Contact Information 1800 060 881 - Beckman Coulter Customer Support Centre