Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-01198-1 |
Product Name/Description |
Combur-Test and ChoiceLine 10 test kits (urine test strips). An in vitro diagnostic medical device (IVD)
ChoiceLine 10 test Material Number: 03504379023
Combur10 Test UX Material Number: 11544373049
Combur10 Test Material Number: 04510089056
Combur9 Test Material Number: 04510054056
All lots affected
ARTG Number: 224003 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
13/09/2016 |
Responsible Entity |
|
Reason/Issue |
The Instructions for Use (IFU) contain an error regarding gram-status of bacteria for the nitrite test. In the Limitations or Accuracy section, gram-negative bacteria have been incorrectly referred to as gram-positive. However, the strip performance is not affected, so the product can still be used as specified. If a physician is not aware that a positive nitrite result is indicative of gram-negative bacteria, but follows the incorrect classification within the IFU, an insufficient class of antibiotic may be chosen for the initial treatment of urinary tract infection (UTI). However, the “gram-classification” of bacteria should be common medical knowledge and use of recommended processes for identification of UTI-causing bacteria (such as urine culture methods) will mitigate risk. Nevertheless, a medical risk for patients due to an insufficient antibiotic chosen initially cannot entirely be excluded. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Roche is advising users of the correct wording for the Instructions for Use (IFU). A revised IFU will be provided to users, and additional information provided with each lot of product until the IFU is corrected. This action has been closed-out on 18/05/2017. |
Contact Information |
02 9860 2357 - Roche Diagnostics Australia |