Type of Product Medical Device
TGA Recall Reference RC-2016-RN-01189-1
Product Name/Description Infant Warmer System (IWS)

Model Numbers: 3000 IWS, 3300 IWS, 3400 IWS and 4400 IWS

ARTG Number: 139290
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 13/09/2016
Responsible Entity GE Healthcare Australia Pty Ltd
Reason/Issue GE Healthcare has recently become aware of a potential safety issue related to loose screws in the “Heater Head” of certain Infant Warmer System (IWS) devices. Hot screws from the “Heater Head” of the IWS could fall onto the bed if the “Heater Head” assembly has been improperly serviced. This could result in thermal injury to a patient. To date, two injuries have been reported as a result of this issue.
Recall Action Recall for Product Correction
Recall Action Instructions GE Healthcare is providing users with interim instructions to check the screws on the "Heater Head" are tight. This will form part of the annual Preventative Maintenance check to ensure the screws are tightened. This action has been closed-out on 23/05/2017.