Type of Product Medical Device
TGA Recall Reference RC-2016-RN-01187-1
Product Name/Description RAPIDPoint 400/405 Systems Measurement Cartridge (blood gas analysers). An in vitro diagnostic medical device (IVD).

RAPIDPoint 400 Measurement Cartridge
Siemens Material Numbers (SMN): 10329756, 10341161

RAPIDPoint 405 Measurement Cartridge
Siemens Material Numbers (SMN): 10844812, 10283222, 10313971, 10310469

Cartridge Serial Numbers: 2610900214 through 2621001325

ARTG Number: 185171
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 9/09/2016
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue Siemens has confirmed an error in the value assignment for the Chloride calibrators in the RAPIDPoint 400 and RAPIDPoint 405 Measurement Cartridge for certain cartridge serial numbers. The maximum % bias observed due to error ranges from -2.3% at 65mmol/L to +5.9% at 132 mmol/L. At extremely high chloride values (>122 mmol/L), a positive bias of 5% or greater may be observed. There have not been any reports of injury due to this issue.
All the other electrolytes (Na+, K+, Ca++) and blood gas parameters are correctly assigned in these measurement cartridges.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is advising users to review the chloride calibrators bias information provided in the recall for product correction letter with their Medical Director.

This action was closed out on the 25/09/2018.
Contact Information 1800 310 300 - Siemens Technical Support Centre