Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-01183-1
Product Name/Description Philips BV25, BV300, BV Libra, BV Endura, BV Pulsera (Fluoroscopic x-ray systems)

System: BV25
Model Number: 718015

System: BV300
Model Numbers: 71815, 71816, 71819

System: BV Libra
Model Numbers: 718022, 718030, 718031

System: BV Endura
Model Numbers: 718071, 718072, 718073, 718074

System: BV Pulsera
Model Number: 718094

ARTG Number: 98556
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 13/09/2016
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue Philips has been notified regarding an issue with the X-Ray "ON" label for systems supplied in South Australia. The colour of the exposure indicator on the control panel is same as the background text. Hence, systems do not meet South Australia's Radiation Protection and Control Regulation.

This issue has been evaluated for radiation hazard and concluded that there is no safety impact to the patient, user or bystander. It is a Regulatory non-compliance issue in South Australia. This issue has been evaluated for potential radiation hazard and concluded that there is no safety impact to the patient, user or bystander.
Recall Action Recall for Product Correction
Recall Action Instructions Philips will be applying a red transparent label over the existing x-ray symbol on the control panel as a permanent correction in the affected systems in South Australia. In the interim, users are advised that there is an indicator lamp on Mobile Viewing Station which glows whenever x-ray exposure happens. The user should use this indication to be able to know when an exposure happens. This action has been closed-out on 24/05/2017.
Contact Information 1800 251 400 - Philips Customer Care Centre