Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-01165-1
Product Name/Description Debriflo Ultrasonic Wound Irrigation Hand-piece

All Lot Numbers

ARTG Number: 265127
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 14/09/2016
Responsible Entity Debriflo Pty Ltd
Reason/Issue It has been found that the current IFU includes an incorrect cleaning and sterilisation procedure. The IFU (Version 14) has been updated to include the new validated cleaning and sterilisation instructions, including the use of a new cleaning solution and wipes, specific dilution directions and contact times, specific lumen flushing instructions and changes to the sterilisation temperature and time.
Recall Action Recall for Product Correction
Recall Action Instructions Debriflo is providing users with a copy of the revised IFU (Version 14) with the new cleaning and sterilisation instructions.

This action has been closed out on 12/08/2019
Contact Information 02 9516 2431 - GSE