Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-01153-1 |
Product Name/Description |
Alaris GP Infusion set with Back Check Valve
Reference Number : 63420EB
Lot Numbers: 1000079, 1000169, 1000177, 1000321
ARTG Number: 125916 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
2/09/2016 |
Responsible Entity |
|
Reason/Issue |
The manufacturer has received an increased rate of reports citing false upstream occlusion alarm of Alaris GP sets with Back Check Valve – 63420EB. These errors have been reported at flow rates below 250mL/hour. There is a risk that a clinician will not be able to start an infusion, or the infusion may be stopped due to an alarm, potentially resulting in a delay in therapy. |
Recall Action |
Recall |
Recall Action Instructions |
BD is advising users to inspect inventory and quarantine any stocks of the affected lot numbers. Affected stock can be returned to BD for credit or replacement.
This action has been closed out on 7/09/2018. |
Contact Information |
1800 833 372 – BD Customer Service |