Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-01149-1
Product Name/Description Coxiella burnetii IFA IgG Kit

Reference Number: PCOBUG
Kit Lot Number: 15C304
Slide Lot Number: 15C403

ARTG Number: 212957

An in vitro diagnostic medical device (IVD)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 8/09/2016
Responsible Entity Abacus ALS Pty Ltd
Reason/Issue Some Coxiella burnetii IFA IgG slides in kits from lot number 15C403 are labelled “Leishmania infantum” instead of “Coxiella burnetii”. The affected slides, while showing the inscription Leishmania infantum, are coated with Coxiella burnetii antigen and have the same features and performance as the slides labelled correctly with the Coxiella burnetii inscription. Of the slides checked by Vircell, only 0.3% present with this labelling anomaly.
Recall Action Recall
Recall Action Instructions Abacus ALS is advising users to cease use of stock with the affected lot number. Abacus will replace affected stock with unaffected stock. A look back of previous results is at the discretion of the laboratory director. This action has been closed out on 13 Jun 2017.
Contact Information 1800 222 287 - Abacus ALS