Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-01149-1 |
Product Name/Description |
Coxiella burnetii IFA IgG Kit
Reference Number: PCOBUG Kit Lot Number: 15C304 Slide Lot Number: 15C403
ARTG Number: 212957
An in vitro diagnostic medical device (IVD) |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
8/09/2016 |
Responsible Entity |
|
Reason/Issue |
Some Coxiella burnetii IFA IgG slides in kits from lot number 15C403 are labelled “Leishmania infantum” instead of “Coxiella burnetii”. The affected slides, while showing the inscription Leishmania infantum, are coated with Coxiella burnetii antigen and have the same features and performance as the slides labelled correctly with the Coxiella burnetii inscription. Of the slides checked by Vircell, only 0.3% present with this labelling anomaly. |
Recall Action |
Recall |
Recall Action Instructions |
Abacus ALS is advising users to cease use of stock with the affected lot number. Abacus will replace affected stock with unaffected stock. A look back of previous results is at the discretion of the laboratory director. This action has been closed out on 13 Jun 2017. |
Contact Information |
1800 222 287 - Abacus ALS |