Type of Product Medical Device
TGA Recall Reference RC-2016-RN-01134-1
Product Name/Description Hamilton Medical C3 Ventilator

Serial Numbers: 1100, 2037

ARTG Number: 194271
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 29/08/2016
Responsible Entity The Trustee for the Taleb Family Trust T/A Taleb Medical and Vetequip Australia
Reason/Issue The manufacturer has identified an issue related to SpO2 monitoring with the use of the Hamilton-C3 ventilator. When measuring SpO2, the software of the Hamilton-C3 can run into a system or CPU overload causing a screen freeze. The alarm buzzer sounds and the device needs to be reset. In case of such a system or CPU overload, the Hamilton-C3 stops the ventilation and sounds the alarm buzzer. However, since the screen freezes, no alarm message is displayed to the operator. There have been no injuries reported as a result of this issue.
Recall Action Recall for Product Correction
Recall Action Instructions All the Hamilton-C3 ventilators have SpO2 option deactivated by default as there are no SpO2 sensors in use in Australia. The new software version 2.0.2 will be installed as a permanent correction. This action has been closed-out on 15/02/2017.