Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-01117-1 |
Product Name/Description |
Dimension & Dimension Vista Urine Opiates Screen assays. In vitro diagnostic medical devices (IVDs).
Assay Name: Dimension Opiates Catalogue Number: DF93A Siemens Material Number: 10444923
Assay Name: Dimension Vista Opiates Catalogue Numbers: K5093 Siemens Material Number: 10445114
All Lots, including all future lots until the IFU is updated
ARTG Number: 182221 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
24/08/2016 |
Responsible Entity |
|
Reason/Issue |
Siemens Healthcare Diagnostics has determined that the concentrations for some of the cross reactants listed in the Specificity Section of the Dimension Opiates assay Instructions For Use and the Cross-reactivity Section of the Dimension Vista Opiates assay Instructions For use are incorrect. Siemens has revised the concentrations, which will enable the correct interpretation on how cross reactants can affect test results. The Instructions For Use for the Opiates assay will be updated with the correct concentrations when Siemens completes their investigation of the issue. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Siemens is advising users to review the revised cross-reactant concentrations for Levallorphan, Levorphanol, Nalorphone and Oxycodone that are provided in the recall correspondence with their Medical Director. The decision to perform a look-back should be at the discretion of the Medical Director. Siemens will be updating the IFU with the revised concentrations once they have completed their investigation of the issue.
This action has been closed out on 19/10/2018 |
Contact Information |
1800 310 300 - Siemens Technical Support Centre |