Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-01115-1 |
Product Name/Description |
Oxoid Antimicrobial Susceptibility Testing Disc - Ceftazidime CAZ10. An in vitro diagnostic medical device (IVD)
Product Code: CT1629B
Lot Number: 1426486
ARTG Number: 208273
|
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
24/08/2016 |
Responsible Entity |
|
Reason/Issue |
Investigation by Oxoid has identified that a single lot of Oxoid Antimicrobial Susceptibility Testing Disc Ceftazidime may contain discs with a reduced concentration of antibiotic, which may impact on performance. Testing with Pseudomonas aeruginosa ATCC 27853 (EUCAST quality control organism) returned lower than specified zones of inhibition. Continued use of this lot could result in quality control failures, delayed results or incorrect results reporting (false resistance). |
Recall Action |
Recall |
Recall Action Instructions |
Thermo Fisher is advising users to inspect their stock for affected units. Affected units are to be removed from further use. The requirement for the review of patient results should be determined by the laboratory director. This action has been closed-out on 11/05/2017. |
Contact Information |
08 8238 9044 - Thermo Fisher Scientific |