Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-01084-1
Product Name/Description QMS Everolimus Immunoassay (used for the therapeutic monitoring of immunosuppressant drugs). An in vitro diagnostic medical device (IVD)

Catalogue Number: 0373852
Lot Number: 72258007
Expiry Date: 31 July 2017

ARTG Number 235822
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 19/08/2016
Responsible Entity Thermo Fisher Scientific
Reason/Issue The manufacturer has determined that the affected lot may yield over-quantitation in patient samples.

The manufacturer has conducted an internal assessment of product performance, and determined that over-recovery of whole blood assay results in the sub-therapeutic range by as much as 50% (for example, a reported assay result of 3.75 ng/mL instead of the actual whole blood concentration of 2.5 ng/mL). Given standardised medication toxicity and close monitoring of patients for markers of efficacy, the risk of developing serious or long-range adverse health consequences due to a single reported erroneous everolimus concentration result is deemed to be negligible.
Recall Action Recall
Recall Action Instructions Thermofisher is advising users to discontinue use and destroy any remaining stocks of the affected batch. A credit note will be issued for the affected product. A review of previously generated results should be considered by the laboratory director and treating physician.

This action has been closed out on 7/09/2018.
Contact Information 02 8817 4279 - Thermo Fisher Scientific