Type of Product Medical Device
TGA Recall Reference RC-2016-RN-01064-1
Product Name/Description Certegra Workstation used in conjunction with Medrad Stellant CT Injection System

Catalogue Number: CWKS DISPLAY

Part Number: PN 3030896

Impacted Workstation Serial Numbers start with KSA051

ARTG Number: 170562
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 11/08/2016
Responsible Entity Imaxeon Pty Ltd
Reason/Issue Medrad has identified that an internal cable connection in the All-in-One computer component of certain Certegra Workstations used in conjunction with the Medrad Stellant CT Injection System may potentially exhibit thermal events such as smoke, sparks or fire. This is a result of degradation over time of the shielding within the connection. If such an event occurs, a delay of procedure, coughing, nausea, difficulty breathing, or first or second degree burns may occur.
There have been no patient or user injuries reported as a result of this situation.
Recall Action Recall
Recall Action Instructions Imaxeon will be arranging for impacted Workstations to be replaced. In the interim, users are advised to take the following actions.

1. While computers are normally warm during use, remain vigilant in checking the back of the Workstation for signs of smoke or other thermal activity while the system is in use. If smoke or other thermal activity is observed, unplug the power cord and discontinue use of your Workstation until it is replaced.

2. Power down the system when you are closing your CT suite for the day or when the CT suite will not be staffed for an extended time and then power it back up prior to your next contrast enhanced CT scan.

Interim Instructions for Use with further details will be provided with the customer letter. This action has been closed-out on 22/02/2017.
Contact Information 02 8845 4999 - Imaxeon Customer Support Team