Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-01046-1 |
Product Name/Description |
1.5 mm HYDRO LeMaitre Valvulotome, 40 cm (intended for the treatment of vascular disorders and particularly for excising or disrupting venous valves)
Reference number: 1010-00
Lot numbers: ELVH1089V, ELVH1091V Expiry date : 2021-03
Lot numbers: ELVH1099V Expiry date : 2021-05
ARTG number: 218706 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
8/08/2016 |
Responsible Entity |
|
Reason/Issue |
The centering hoops keep the head of the HYDRO LeMaitre Valvulotome centered in the vessel and prevents the valve-cutting blades from damaging the vessel wall. There have been reported issues of hoops failing to close when the device was actuated. In some cases, this issue has been discovered in-use. While no adverse events have been reported, there is the possibility that a malfunctioning device could damage the vessel upon withdrawal. |
Recall Action |
Recall |
Recall Action Instructions |
Emergo is advising users to locate, quarantine and return all affected product. Returned devices will be replaced with unaffected devices. This action has been closed-out on 02/03/2017. |
Contact Information |
02 9006 1662 - Emergo Australia |