Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-01014-1 |
Product Name/Description |
Ceiling Bracket, Arch 71 White (used for installing Liko patient lifts to wooden and concrete ceilings)
Product code: 3102071
ARTG number: 161863 |
Recall Action Level |
Wholesale |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
15/08/2016 |
Responsible Entity |
|
Reason/Issue |
Hill-Rom has received one complaint where the Ceiling Bracket 71 has been shipped with an increased size of the center hole.
A potentially hazardous situation exists if the Ceiling Bracket 71 with a larger center hole is used for a concrete ceiling installation since a M8 screw and washer will be used. Analysis has shown that if the bracket is mounted with a M8 screw and washer in a concrete ceiling Ultra System Installation (maximum load of 400 – 500 kg) there is a safety hazard associated with the installation that could potentially cause the system to fall. |
Recall Action |
Recall |
Recall Action Instructions |
Hill-Rom is advising the distributor to inspect stock and quarantine the affected brackets for return. All the affected stock is with the distributor. This action has been closed-out on 10/05/2017. |
Contact Information |
1800 445 576 - Hill-Rom Technical Support |