Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00996-1
Product Name/Description Sterrad 100NX system (Sterilisation system)

Part Numbers: 10104, 10104002, 10104003, 10104004

ARTG Number: 123603
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 26/07/2016
Responsible Entity Johnson & Johnson Medical Pty Ltd
Reason/Issue Johnson & Johnson Medical (JJM) is advising that the Sterrad 100NX system requires a CTick label (RFID) be applied to comply with the Australian Communications and Media Authority (ACMA) spectrum licensing and CTick requirements.
Recall Action Recall for Product Correction
Recall Action Instructions Johnson & Johnson Medical will be attending all sites to place the required label on the affected devices. This action has been closed-out on 13/07/2017.
Contact Information 02 9815 3765 - Johnson & Johnson Medical