Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00975-1 |
Product Name/Description |
Molift Air 205 and 300 ceiling hoist (a powered lifting device)
Molift Air 205 Model Numbers: 2501205, 2501210, 2501215, 2501220
Molift Air 300 Model Numbers: 2501300, 2501310, 2501315, 2501320
Distributed between May 2013 and January 2016
ARTG Number: 229696 |
Recall Action Level |
Consumer |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
25/08/2016 |
Responsible Entity |
|
Reason/Issue |
The manufacturer has received a total of 4 reports where the bolt on the suspension coupling on Molift Air has become loose in the coupling. All reports came after technical testing and inspection. To date, no reports have been received regarding patient injury resulting from this issue. However, there is a potential risk for harm or injury if the bolt inside the coupling comes loose from the housing. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Patient Handling will arrange to upgrade the device with a retainer safety ring that will prevent the bolt within the suspension coupling coming loose. Qualified service personnel will install the upgrade kits on all affected systems. Consumers are advised to contact the place of purchase to organise a qualified service technician to install the upgrade kit. For more details, please see https://www.tga.gov.au/alert/molift-air-ceiling-hoist . This action has been closed-out on 21/03/2018. |
Contact Information |
1300 137 875 - Patient Handling Pty Ltd |