Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00930-1 |
Product Name/Description |
Savion ED Trolley (mobile stretcher)
Model Number: ES-711X
Serial Number: 20150811XXXX, 201508170001 and 201508170002
ARTG Number: 225146 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
12/07/2016 |
Responsible Entity |
|
Reason/Issue |
It has been determined that the a number of trolleys with the affected serial numbers have been supplied to ArjoHuntleigh with the incorrect TBB6 battery. This may result in a burning smell when connected to mains power. There have been no reports of injuries as a result of this issue. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
ArjoHuntleigh has advised users to remove the stretchers from use. The batteries will be replaced by unaffected stock as a permanent correction. This action has been closed-out on 03/08/2016. |