Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00930-1
Product Name/Description Savion ED Trolley (mobile stretcher)

Model Number: ES-711X

Serial Number: 20150811XXXX, 201508170001 and 201508170002

ARTG Number: 225146
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 12/07/2016
Responsible Entity Huntleigh Healthcare Pty Ltd
Reason/Issue It has been determined that the a number of trolleys with the affected serial numbers have been supplied to ArjoHuntleigh with the incorrect TBB6 battery. This may result in a burning smell when connected to mains power. There have been no reports of injuries as a result of this issue.
Recall Action Recall for Product Correction
Recall Action Instructions ArjoHuntleigh has advised users to remove the stretchers from use. The batteries will be replaced by unaffected stock as a permanent correction. This action has been closed-out on 03/08/2016.