Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00928-1 |
Product Name/Description |
Avenir Müller Stem 6 Lateral uncemented and Avenir Müller Stem 4 Standard uncemented (used as the component in total hip arthoplasty)
Avenir Müller Stem 6 Lateral uncemented Reference Number: 01.06010.106 Lot Number: 4023094
Avenir Müller Stem 4 Standard uncemented Reference Number: 01.06010.004 Lot Number: 4022860
ARTG Number: 216175 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
12/07/2016 |
Responsible Entity |
|
Reason/Issue |
The Avenir Müller Stem 6 lateral uncemented may have been placed in the packaging of the Avenir Müller Stem 4 standard uncemented and vice versa.
The surgeon may have used a standard stem instead of a lateral stem. For the standard stem the offset is 6mm smaller and therefore an increased laxity of the hip might result. This could conduct to a risk of luxation, pain or revision surgery. |
Recall Action |
Hazard Alert |
Recall Action Instructions |
Zimmer is advising users that use of a standard stem instead of a lateral stem may result in an increased laxity of the hip, and may result in pain or revision surgery. This action has been closed-out on 17/02/2017. |
Contact Information |
02 9483 5426 - Zimmer Biomet |