Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00923-1
Product Name/Description Frazier Suction instrument (used for precision suction in spinal or neurological surgeries)

Product Codes: 309.02.006, 309.02.008, 309.02.008L, 309.02.010, 309.02.010L, 309.02.012, 309.02.012L

All lots affected

ARTG Number: 172620
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 12/07/2016
Responsible Entity ConvaTec Australia Pty Ltd
Reason/Issue ConvaTec (Australia) Pty Ltd is issuing an Instruction for Use (IFU) for their Unomedical brand Frazier Suction Instrument which currently is supplied without an IFU. This action is being taken in response to complaints from the users in relation to breaking of the instrument at the junction of the handle and the metal tubing when pressure is applied at the tip.

If the issue were to occur, it would result in a loss of suction and loss of visibility in the surgical field. There have been no reports of injuries as a result of this issue.
Recall Action Recall for Product Correction
Recall Action Instructions The manufacturer has decided to issue an IFU to instruct the users on how to use the product appropriately to avoid any breakage of the instrument. ConvaTec is providing hospitals with the IFU along with the customer letter. This action has been closed-out on 07/02/2017.