Reason/Issue |
Stryker has received complaints from customers about the device dissociating or breaking during use. An analysis indicated that the press fit specifications between the dowel pin and the mating hole in the shaft were not being achieved by the supplier, potentially resulting in disassociation from the instrument. There is a risk that the handle components, including a dowel pin and a locking pin, may potentially disassociate and fall into the wound intraoperatively, necessitating retrieval. There have been no injuries reported as a result of this issue. |