Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00910-1
Product Name/Description Triathlon Modular Handle
(a reusable surgical instrument used in total and partial knee replacements)

Item Number: 6541-4-808

All Lot Numbers

ARTG Number: 140892
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 8/07/2016
Responsible Entity Stryker Australia Pty Ltd
Reason/Issue Stryker has received complaints from customers about the device dissociating or breaking during use. An analysis indicated that the press fit specifications between the dowel pin and the mating hole in the shaft were not being achieved by the supplier, potentially resulting in disassociation from the instrument. There is a risk that the handle components, including a dowel pin and a locking pin, may potentially disassociate and fall into the wound intraoperatively, necessitating retrieval. There have been no injuries reported as a result of this issue.
Recall Action Recall
Recall Action Instructions Stryker is advising users to inspect inventory and remove the affected devices from the procedure trays. The trays can continue to be used once the affected items have been removed. The affected devices can be returned to Stryker.
This action has been closed-out on 23/03/2018.