Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00892-1
Product Name/Description Quickie Q7 Manual Wheelchair

Model Number: EIR4
Serial Number Range: R4-007239 to R4-022394
Manufactured between 1 September 2009 and 27 November 2013

ARTG Number: 100167
Recall Action Level Consumer
Recall Action Classification Class II
Recall Action Commencement Date 7/07/2016
Responsible Entity Sunrise Medical Pty Ltd
Reason/Issue It has been identified that over time, and under some usage conditions, the backrest brackets may break and detach from the chair frame. If a backrest bracket breaks and goes unnoticed by the user or care giver, there is a risk that the occupant could fall out of the seat and sustain injury to the head or lower back. The expected failure rate of the issue is 1.75%. There have been four reported injuries as a result of users unexpectedly falling from the seats. If a backrest bracket breaks and goes unnoticed by the user or caregiver, there is a risk for the occupant to fall out of the seat and sustain an injury to the head or lower back.
Recall Action Recall for Product Correction
Recall Action Instructions Consumers will be contacted by the dealer who supplied the device to arrange replacement of affected backrest brackets. Sunrise Medical is providing dealers with a replacement bracket kit to be installed on all affected wheelchairs. Wheelchairs manufactured after 27 November 2013 are unaffected by this issue. For more details, please see https://www.tga.gov.au/alert/quickie-q7-manual-wheelchair .

This action was closed out on the 24/09/2018.