Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00870-1
Product Name/Description Merge PACS with software versions 6.0MR2 and greater up to and including 6.6.2.2. Merge PACS versions 7.0 and greater up to and including 7.0.2
(a picture archiving and communication system)

ARTG Number: 191068
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 30/06/2016
Responsible Entity Emergo Asia Pacific Pty Ltd T/a Emergo Australia
Reason/Issue Unviewed images may not display after the last Hanging Protocol (HP) step when the user preference setting "Show unviewed images after the last hanging Protocol (MG only)" is selected. This can occur if the last HP step was skipped. This is most commonly encountered if there are additional views created for the study.
This may impact diagnosis when images are unavailable for viewing, however there have been no reports of injuries as a result of this issue.
Recall Action Recall for Product Correction
Recall Action Instructions Emergo is advising users to not have the red-x icon available in the reading physician's toolbars to close the study. This will trigger a warning message that there are unviewed images available.
Emergo will be implementing a software update as a permanent correction. This action has been closed-out on 27/02/2017.