Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00856-1 |
Product Name/Description |
Torque Limiting Adapter for use in orthopaedic and trauma surgical procedures (Re-useable surgical instrument)
Part Number: 00-2360-080-00
All lot numbers affected
ARTG Number: 260719 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
27/06/2016 |
Responsible Entity |
|
Reason/Issue |
The manufacturer, Greatbatch Medical identified through laboratory testing that the required sterility assurance level (SAL) of 10-6 is not achieved on Torque Limiting Adapter when subjected to steam sterilisation cycles identified in the Instructions For Use (IFU) provided with the device. Torque Limiting Adapter is supplied non-sterile and requires cleaning and steam sterilisation prior to use and subsequent re-use. |
Recall Action |
Recall |
Recall Action Instructions |
Zimmer Biomet is advising users to immediately locate and remove the identified device. Zimmer Biomet will remove the affected product from the facility. |
Contact Information |
02 9483 5426 - Zimmer Biomet |