Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00837-1
Product Name/Description ADVIA Centaur XPT System. An in-vitro diagnostic medical device (IVD)

Siemens Material Number: 10711433

ARTG Number: 175890
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 24/06/2016
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue Siemens has determined that certain waste reservoirs in ADVIA Centaur XPT System were manufactured with walls that are below the nominal specification, potentially causing the following problem when the reservoir is under vacuum:

- The seam at the bottom of certain reservoirs could crack slightly and have a slight leak.
- The fitting at the bottom of the reservoir could exhibit some leaks at the interface between the reservoir and the fitting.

If a leak occurs due to a crack in the bottom of the reservoir, foaming may occur in the glass jar in the bulk fluid drawer and cause the instrument to stop with vacuum low error # 01 600 13 24.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens are advising users to avoid any potential leak onto the floor by performing the following action during the weekly maintenance:

- Verify that no fluid has accumulated under the Waste reservoir.
- Verify that the glass jar in the bulk fluid drawer is empty and is not foaming.

Siemens will replace the defective reservoir.

This action has been closed out on the 16/10/2018.
Contact Information 1800 310 300 - Siemens Technical Support Centre