Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00833-1 |
Product Name/Description |
cobas 6800 system - cobas MPX, 96 tests and cobas MPX, 480 tests. An in-vitro diagnostic medical device (IVD).
cobas MPX, 96 tests Material Number: 06997708190 All Lot Numbers affected
cobas MPX, 480 tests Material Number: 06997716190 All Lot Numbers affected
ARTG Number: 275469 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
27/06/2016 |
Responsible Entity |
|
Reason/Issue |
Roche Diagnostics Australia is informing users that in extremely rare cases, sample under-aspiration may occur when using the cobas 6800 system with sample input volumes of less than 450 µL. This issue will only affect the cobas MPX test when used with cadaveric donors, since the minimum sample volume required for cadaveric donors is 300 µL (for living donors, the minimum sample volume required is 1,000 µL).
If bubbles or foam are present in the sample tube, this may lead to false sample liquid level detection and under-aspiration of the sample. This could potentially cause false negative / under-quantitated results that are not flagged. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Roche is advising users to ensure good sample preparation technique to avoid the formation of foam, clots, or bubbles in all samples, as described within the instrument safety manual. A look-back at previous cadaveric results is also recommended. Roche will be implementing a software correction as a permanent fix.. This action has been closed-out on 22/03/2018. |
Contact Information |
02 9860 2313 - Roche Diagnostics |