Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00814-1
Product Name/Description QIAsymphony DSP DNA Mini Kits. An in vitro diagnostic medical device (IVD).

Catalogue number: 937236

Lot number: 154023773

Reagent cartridge lot number: 154022318

ARTG number: 211651
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 21/06/2016
Responsible Entity Qiagen Pty Ltd
Reason/Issue Some reagent cartridges (REAG CART) in this kit lot contain a reagent cartridge holder with an incorrect shape. One “fixing projection” between Well 1 and the bead trough is significantly larger and protruding. This could lead to interrupted runs or incorrect sample processing.
Recall Action Recall
Recall Action Instructions QIAGEN is advising users to quarantine any remaining stocks of the affected product. Product will be inspected and replaced if affected.
QIAGEN is recommending a review of previously generated results. This action has been closed-out on 20/02/2017.
Contact Information 03 9840 9800 - QIAGEN Technical Services Department