Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00807-1
Product Name/Description Dimension Creatinine (CRE2) Flex reagent cartridge (Used with Dimension clinical chemistry system). An in vitro diagnostic medical device (IVD)

Catalogue Number: DF33B

Siemens Material Number: 10872079

All lot numbers affected

ARTG Number: 181689
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 16/06/2016
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue Siemens Healthcare Diagnostics has identified that Dimension Creatinine (CRE2) Assay exhibits a negative bias at the low end of the urine Analytical Measurement Range (AMR). The Limit of Quantitation (LoQ) claim 442 µmol/L for urine samples is not met. This issue affects all future lots of CRE2 until a solution is implemented. The serum/plasma CRE2 AMR is not affected by this issue.

The bias observed for urine creatinine samples < 1149 ìmol/L would not impact clinical interpretation of adult eGFR calculations or clinical interpretation of laboratory tests utilising creatinine as a correction factor, such as Urine Albumin/Creatinine ratio and/or Urine Protein/Creatinine ratio.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is advising users to review the maximum observed bias for low urine concentrations. It is recommended that urine CRE2 values that are less than 1149umol/L are reported as <1149umol/L rather than an actual numeric results. Siemens is providing work around instructions for users to change the lower assay range parameters to 1149umol/L. Siemens is not recommending a review of previously generated results, however a review is at the discretion of the laboratory.

This action has been closed out on 28/09/2018
Contact Information 1800 310 300 - Siemens Technical Support Centre