Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00803-1
Product Name/Description V-PRO 1, V-PRO 1 Plus, V-PRO maX Low Temperature Sterilisation Systems

Manufactured between 2 January 2014 and 2 May 2016

ARTG Number: 154905
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 20/06/2016
Responsible Entity Device Technologies Australia Pty Ltd
Reason/Issue STERIS has identified that V-PRO 1, V-PRO 1 Plus, and V-PRO maX Sterilisers manufactured after 1 January 2014 may not display the correct date as of
29 February 2016. While the V-PRO units display an incorrect date, the steriliser functions and cycle operations continue to operate properly. If users do not manually change the system date of the V-PRO unit, the unit’s display and cycle printout will contain an incorrect date.
Recall Action Recall for Product Correction
Recall Action Instructions Device Technologies is advising users that a service technician will be upgrading all affected systems. Users are advised that the system date can be amended using the instructions found in the Operator's Manual. The system can continue to be used as normal until corrected. This action has been closed-out on 11/05/2017.
Contact Information 02 8244 3721 - Device Technologies