Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00771-1 |
Product Name/Description |
Sirona Dac Universal Combination Steriliser
Serial Number Range: 104000-199999, 202000-999999, B14000-B99999, B104000-B199999, B202000-B999999
ARTG Number: 197032 |
Recall Action Level |
Retail |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
20/06/2016 |
Responsible Entity |
|
Reason/Issue |
Dac Universal MKIII units may have been repaired using spare parts intended for MKII units. There is the possibility of 12mm screws being used as a replacement for the original 16mm screws in the lid holder assembly. In the case, where the incorrect screws are fitted the lid could possibly become unmounted from the unit during its operation. There is a risk that people standing nearby could be harmed if hit by a released lid. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Sirona is advising users to discontinue use of the affected machines until a service technician has inspected and corrected the machine, where required. Users are reminded to continue to manually clean and lubricate the instruments using T1 spray. This action has been closed out on 15 June 2017. |
Contact Information |
02 9496 5100 - Sirona |