Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00746-1
Product Name/Description LumaDent Light Battery Charger

All model numbers affected EXCEPT SW-084120

ARTG Number: 207981
Recall Action Level Retail
Recall Action Classification Class II
Recall Action Commencement Date 3/06/2016
Responsible Entity Carl Zeiss Pty Ltd
Reason/Issue Carl Zeiss Pty Ltd, following advice from LumaDent is replacing the battery charger of the LumaDent Light. This applies to all charger model numbers, except model number SW-084120. Model number SW-084120, which is the only correct model for all LumaDent batteries BP500 (V3.1, V3.2 and V3.3), is not affected by this recall.

The affected battery chargers have not been demonstrated to comply with Australian and New Zealand Electrical Safety Standards thus may have the potential to overcharge the batteries causing them to overheat and catch fire. There have been no reports of injuries as a result of this issue.
Recall Action Recall
Recall Action Instructions Carl Zeiss are advising users to inspect all battery chargers and quarantine any batteries which are not model number SW-084120. Carl Zeiss will replace any affected stock. This action has been closed-out on 02/03/2017.
Contact Information 1300 365 470 - Carl Zeiss