Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00745-1
Product Name/Description Compella Bariatric Bed System, Controllers and Air Surfaces

Model Number: P7800A

Compella Controllers
Product Number: P7810A
Manufactured between 25 March 2015 to 11 December 2015

Compella Air Surfaces
Product Number: P642A
Manufactured between 25 March 2015 to 25 September 2015

ARTG Numbers: 239273, 239274
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 8/06/2016
Responsible Entity Hill-Rom Pty Ltd
Reason/Issue Hill-Rom has become aware of a potential issue with some Compella Air Supply Units and Air Surfaces. The Compella Air Supply Unit has been prone to inadequate pressure, which prevents the surface from achieving required settings for weight-based pressure redistribution. This also prevents the low air loss feature from operating. In addition, within the Low Air Loss surface, a hose in the foot section is prone to kinking, which can impact airflow to the topper and prevent the proper adjustment of the air bladders to relieve pressure.
Recall Action Recall for Product Correction
Recall Action Instructions A Hill-Rom representative will be contacting affected facilities to make arrangements for updating the affected capital Compella Air Supply Units and air surfaces. In the interim, Hill-Rom recommends customers to continue to use the product as they normally would. The mattress system will provide an alarm in the event the Compella Air Supply Unit or Air Surface does not operate properly. This action has been closed-out on 18/05/2017.