Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00729-1
Product Name/Description AIRVO 2 / myAIRVO 2 Humidifier

AIRVO 2 Humidifier
Model Number: PT101XX

myAIRVO 2 Humidifier
Model Number: PT100XX

Affected Serial Numbers: 120521YYYYYY – 160605YYYYYY

ARTG Number: 177298
Recall Action Level Retail
Recall Action Classification Class III
Recall Action Commencement Date 8/06/2016
Responsible Entity Fisher & Paykel Healthcare Pty Ltd
Reason/Issue Fisher & Paykel has updated the AIRVO 2 / myAIRVO 2 User Instructions to include a check that the speaker is audible prior to each patient use in order to ensure the speaker is functioning.
Recall Action Recall for Product Correction
Recall Action Instructions Fisher & Paykel are advising users that the AIRVO 2 / myAIRVO 2 User Instructions have been updated to include a warning for the user to check speaker functionality before each patient use, using the steps in the User Instructions. First turn the unit on and then remove the heated breathing tube. The user should see the “check tube” visual signal and hear the speaker signal.
If either signal is absent, users are advised not to use the unit and contact their Fisher & Paykel representative to arrange for a replacement device.

This action has been closed out on the 12/10/2018.
Contact Information 03 9879 5022 - Fisher & Paykel Healthcare