Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00729-1 |
Product Name/Description |
AIRVO 2 / myAIRVO 2 Humidifier
AIRVO 2 Humidifier Model Number: PT101XX
myAIRVO 2 Humidifier Model Number: PT100XX
Affected Serial Numbers: 120521YYYYYY – 160605YYYYYY
ARTG Number: 177298 |
Recall Action Level |
Retail |
Recall Action Classification |
Class III |
Recall Action Commencement Date |
8/06/2016 |
Responsible Entity |
|
Reason/Issue |
Fisher & Paykel has updated the AIRVO 2 / myAIRVO 2 User Instructions to include a check that the speaker is audible prior to each patient use in order to ensure the speaker is functioning. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Fisher & Paykel are advising users that the AIRVO 2 / myAIRVO 2 User Instructions have been updated to include a warning for the user to check speaker functionality before each patient use, using the steps in the User Instructions. First turn the unit on and then remove the heated breathing tube. The user should see the “check tube” visual signal and hear the speaker signal. If either signal is absent, users are advised not to use the unit and contact their Fisher & Paykel representative to arrange for a replacement device.
This action has been closed out on the 12/10/2018. |
Contact Information |
03 9879 5022 - Fisher & Paykel Healthcare |