TGA Recall Reference RC-2016-RN-00721-1
Product Name/Description Drager Ventstar Paediatric Circuit, used in conjunction with the Dräger Oxylog 3000plus Transport Ventilator

ARTG Number: 131796
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 27/05/2016
Responsible Entity Draeger Medical Australia Pty Ltd
Reason/Issue During the course of routine internal testing Drager detected isolated leakages at the check valve (one way valve of the inspiration branch) of the above mentioned disposable paediatric ventilation circuit. Leakage at the check valve can result in a patient's exhaled gas entering into the breathing circuit, which could lead to the rebreathing of the exhaled gas with subsequent reduced oxygen concentration for the patient. This leakage is not detected during the ventilator operational readiness check. To date, Drager has not received any complaints associated with this issue.

Further detailed investigations have shown that the check valve functions properly with no leak observed at positive end-expiratory pressures above 5 mbar I cmH20. The problem only occurs at PEEP values below 5 mbar I cmH20.
Recall Action Recall
Recall Action Instructions Drager urgently recommends that users immediately inspect any stock and dispose of any product (part number 5704964) with a date of manufacture up to and including 03-2016. A Free of charge replacement product can be obtained by following the instructions in the customer letter.
This action has been closed-out on 22/03/2018.