Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00680-1 |
Product Name/Description |
Philips Brilliance iCT/iCT SP with software version 4.1.3, 4.1.5
ARTG Number: 98868 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
26/05/2016 |
Responsible Entity |
|
Reason/Issue |
Philips Healthcare became aware of certain issues with the affected products through testing & customer feedback. The identified issues include; - IMR Cardiac - Irregular Appearance of Contrast in Vessels; - Missing Image Annotations in CCT - Halo Artefacts - Incorrect Z Annotation on Plan Box for Coronal/Sagittal Surviews - Post Injection Delay Between Timed Scans may be Inaccurate - Unintended Change of Acquisition Timing on Scan Ruler - Bolus Tracker does not Trigger as Expected - Cardiac Step & Shoot stops after Arrhythmia detected - Tracker Scan Halted with FDOM - Error when Auto ROI Placement Outside Patient Anatomy - No "GO" when HR outside acceptable range - Unplanned Results During Multi-Phase Pulmo Series, may Cause a System Crash - Dynamic Myocardial Perfusion Acquisition Halted After 1st Shot - Double Clicking ECG Phase Bar Crashes Exam Application - Incorrect Phase Tolerance for Cardiac S&S & - Ring Artefacts on High Resolution Images, iDose4 & IMR |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Philips is providing work around instructions for users to follow as an interim measure. A software upgrade will be implemented as a permanent correction. This action has been closed-out on 24/05/2017. |
Contact Information |
1800 251 400 - Philips Customer Care Centre |