Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00680-1
Product Name/Description Philips Brilliance iCT/iCT SP with software version 4.1.3, 4.1.5

ARTG Number: 98868
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 26/05/2016
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue Philips Healthcare became aware of certain issues with the affected products through testing & customer feedback. The identified issues include;
- IMR Cardiac - Irregular Appearance of Contrast in Vessels;
- Missing Image Annotations in CCT
- Halo Artefacts
- Incorrect Z Annotation on Plan Box for Coronal/Sagittal Surviews
- Post Injection Delay Between Timed Scans may be Inaccurate
- Unintended Change of Acquisition Timing on Scan Ruler
- Bolus Tracker does not Trigger as Expected
- Cardiac Step & Shoot stops after Arrhythmia detected
- Tracker Scan Halted with FDOM
- Error when Auto ROI Placement Outside Patient Anatomy
- No "GO" when HR outside acceptable range
- Unplanned Results During Multi-Phase Pulmo Series, may Cause a System Crash
- Dynamic Myocardial Perfusion Acquisition Halted After 1st Shot
- Double Clicking ECG Phase Bar Crashes Exam Application
- Incorrect Phase Tolerance for Cardiac S&S &
- Ring Artefacts on High Resolution Images, iDose4 & IMR
Recall Action Recall for Product Correction
Recall Action Instructions Philips is providing work around instructions for users to follow as an interim measure. A software upgrade will be implemented as a permanent correction. This action has been closed-out on 24/05/2017.
Contact Information 1800 251 400 - Philips Customer Care Centre