Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00678-1 |
Product Name/Description |
Self Tapping Bone Screws, 6.5 x 20 and 6.5 x 25
Self Tapping Bone Screw 6.5 x 20 Part Number: 00-6250-065-20 Lot Number: 63301504
Self Tapping Bone Screw, 6.5 x 25 Part Number: 00-6624-065-25 Lot Number: 63276975
ARTG Number: 169898 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
25/05/2016 |
Responsible Entity |
|
Reason/Issue |
Zimmer Inc. has become aware of specific lots of sterile Hip and Trauma Implant Screws manufactured between February 2016 and March 2016. These screws are packaged with a double barrier system, two plastic trays and two Tyvek lids, to provide sterile integrity. Process monitoring conducted as part of the standard packaging process identified that in some instances, a small hole may be present in the inner tray, which were found to be caused by a single tool used by the tray supplier in the manufacturing process and estimated to be present in less than 10% of the affected devices. The outer tray is not affected and the device’s sterile integrity remains until the outer tray is opened. |
Recall Action |
Hazard Alert |
Recall Action Instructions |
Zimmer is advising surgeons to review the notification and ensure affected personnel are aware of the contents. Patient follow up is at the discretion of the treating surgeon. This action has been closed-out on 27/01/2017. |
Contact Information |
02 9483 5426 - Zimmer Biomet |