Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00638-1
Product Name/Description DigitalDiagnost

Affected Products: DigitalDiagnost Release 4.0.4, 4.1.2, 4.1.3

ARTG Number: 117662
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 13/05/2016
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue A firmware anomaly inside the detector (4343RC) can cause the detector to appear ready for acquisition on very short time intervals while it actual fact it is not ready, resulting in incorrectly acquired X-ray images.

If a corrupted image cannot be used for its clinical purpose, it may be necessary to retake the image, leading to additional exposure for the patient.
Recall Action Recall for Product Correction
Recall Action Instructions Philips plans to install a software upgrade in affected systems, which will eliminate this issue. A Philips Service Engineer will contact the affected customers when the Field Action Kit is available to be implemented. This action has been closed-out on 16/05/2017
Contact Information 1800 251 400 - Philips Customer Care Centre