TGA Recall Reference RC-2016-RN-00618-1
Product Name/Description AVOXimeter System (a whole blood oximeter used during heart catheterisation procedures)

Product Code: AVOX1000E

Serial Numbers: 5696-5725, 5816-5977

ARTG Number: 262420
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 12/05/2016
Responsible Entity Zoll Medical Australia Pty Ltd
Reason/Issue The manufacturer, Accriva Diagnostics has confirmed that some AVOXimeter 1000E instruments shipped to customers from July 2015 and April 2016 may have been sent to customers with the diagnostic mode enabled.
With the diagnostic mode enabled, users could mistakenly unlock features in the instrument software through a specific sequence of keystrokes. If these features are altered, there is a potential for erroneous results to be displayed. To date Accriva has not received any reports of injuries or harm to users due to this issue. Instruments in diagnostic mode perform per Accriva’s specifications and all data generation is equivalent to those instruments properly set to user mode.
Recall Action Recall for Product Correction
Recall Action Instructions Zoll Medical is providing instructions for users to be able to identify the affected instruments. In the case where an affected instrument is identified, users should contact Zoll Technical Services to correct the settings. This action has been closed-out on 03/02/2017.