Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00599-1 |
Product Name/Description |
EliA IgG Conjugate. An in vitro diagnostic medical device (IVD).
EliA IgG Conjugate 50 Article Number: 83-1017-01
EliA IgG Conjugate 200 Article Number: 83-1018-01
EliA IgG Conjugate Article Number: 83-1005-01
EliA IgG Conjugate Article Number: 83-1002-01
Multiple Lots and Kits affected
ARTG Number: 229585 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class III |
Recall Action Commencement Date |
27/05/2016 |
Responsible Entity |
|
Reason/Issue |
Investigation of complaints received has concluded that when using EliA CCP Well (14-5515-01) in combination with EliA IgG Conjugate (83-1017-01, 83-1005-01, 83-1002-01 or 83-1018-01), anti-CCP-signals can be elevated for the listed conjugate lots.
The positive bias of the EliA CCP test will change a small percentage of test results from negative to equivocal and equivocal or high negative to low positive.
Testing on clinical samples of both confirmed Rheumatoid Arthritis (RA) patient sera, healthy blood donors and other disease controls resulted in < 1% of the samples changing from negative to positive, according to current cut-off ranges (total study includes 472 samples). The results of the assay evaluation studies performed confirm the variance to be within current clinical performance, and it is concluded that the clinical impact of the observed bias is negligible. |
Recall Action |
Recall |
Recall Action Instructions |
The sponsor is advising customers to stop using EliA CCP Well (14-5515-01) in combination with EliA IgG Conjugate (83-1017-01, 83-1005-01, 83-1002-01 or 83-1018-01) of the affected lots as stated. The sponsor will provide unaffected EliA IgG conjugate lots as soon as they become available. This action has been closed-out on 15/02/2017. |
Contact Information |
1800 222 287 - Abacus ALS |