Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00563-1 |
Product Name/Description |
Novocastra Liquid Mouse Monoclonal Antibody Calretinin. An in vitro diagnostic medical device (IVD)
Product Code: 1ml NCL-L-CALRET-566 Lot Numbers: 6025204, 6029060, 6036060
Only to be used in combination with:
Diluent Lot Numbers
AR9352 SLBL4196V, SLBL8154V, SLBM1997V, SLBM3199V
AR9352-CN SLBL8154V SLBM1997V
RE7133-CE 6035000
RE7133-CN 6035115
ARTG Number: 178442
|
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
29/04/2016 |
Responsible Entity |
|
Reason/Issue |
In the event that the product is used with diluents other than those listed, it may not provide adequate staining. When used according to the Instructions for Use with appropriate positive controls, the failure to adequately stain would be apparent to the user on the control tissue and on the patient tissue. In these instances, the absence of adequate staining of tissue is likely to result in a delay in obtaining the results of IHC / ISH staining and, in a worst-case scenario, could cause a delay in the diagnosis or classification of a neoplasm.
|
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Leica is advising users to only use Novocastra Liquid Mouse Monoclonal Antibody Calretinin with the specified diluents. There is no need for retrospective review of tissue staining if appropriate positive controls were properly used and interpreted during the immuno / in situ staining. This action has been closed-out on 25/01/2017. |
Contact Information |
03 9211 7535 - Leica Biosystems |