Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00560-1
Product Name/Description Maxi Air Blower Unit - Model 230V
(a device designed as a stretcher that is intended for use together with a patient transfer lifting for lifting a patient who is moveable only in a horizontal lying position)

Serial Numbers: S0112300, S0112299, S0102544

ARTG Number: 215885
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 29/04/2016
Responsible Entity Huntleigh Healthcare Pty Ltd
Reason/Issue The plastic enclosure used in the Maxi Air supply motor for models 230V is not fire retardant. This can be seen by the "arrows" on the moulded part and was confirmed by an ad-hoc flammability test. This may increase the risk of a fire spreading.
Recall Action Recall for Product Correction
Recall Action Instructions An ArjoHuntleigh representative will attend the affected hospital to replace the affected blower unit.

This action has been closed out on 02/11/2018
Contact Information 08 9314 4117 - ArjoHuntleigh